Implantable Pacemaker
What is a pacemaker?
- A pacemaker (artificial pacemaker) is a medical device which uses electrical impulses, delivered by electrodes contracting the heart muscles, to regulate the beating of the heart.
- The primary purpose of a pacemaker is to maintain an adequate heart rate, either:
- Because the hearts natural pacemaker (the SA node) is not fast enough.
- Because there is a block in the hearts electrical conduction system.
- In modern pacemakers, cardiologist can externally program and select the optimum pacing modes for individual patients.
- Some combine a pacemaker and defibrillator in a single implantable device. It could have multiple electrodes stimulating differing positions within the heart.
Pacemaker are usually used:
- When a patient has a permanent or temporary slower than normal impulse formation.
- When a symptomatic AV or ventricular conduction disturbance.
- To control some tachydysrhythmias that do not response to medication.
Pacing Systems
A Unipolar Pacing System Contains a Lead with Only One Electrode Within the Heart
In this system, the Impulse:
- Flows through the tip electrode (cathode)
- Stimulates the heart
- Returns through body fluid and tissue to the IPG (anode)
A Bipolar Pacing System Contains a Lead with Two Electrodes Within the Heart
In this system, the Impulse:
- Flows through the tip electrode located at the end of the lead wire (cathode)
- Stimulates the heart
- Returns to the ring electrode above the lead tip (anode)
- Positions I and II indicate which chambers are to be paced and sensed.
- Position III relates to whether a pacing stimulus is either triggered or inhibited in response to a sensed cardiac depolarization.
- Position IV is used to determine whether the pacing rate can be automatically modulated (adjusted) by the pacemaker. The pacemaker can be configured for adaptive pacing to detect physiological parameters related to respiratory, body movement, temperature, oxygen saturation, and pH in the blood, in response to exercise such that the heart rate can be increased in order to cope with the extra physiological demands.
- Position V indicates whether multisite pacing is available in one or more of the chambers (Cardiac Resynchronization Therapy).
Examples of pacemaker settings:
- AOO—Atrial asynchronous pacemaker.
- Pace in atria, no chamber sensing. Atrial pacing at a fixed rate that is independent of the inherent heart rate.
- VOO—Ventricle asynchronous pacemaker.
- Pace in ventricle, no chamber sensing.
- Ventricular pacing at a fixed rate that is independent of the inherent heart rate.
- AAI—Atrial single chamber synchronous pacemaker.
- Pace in atrium and sense in atrium.
- Atrium pacing inhibited in response to atrium activity.
- VVI—Ventricular single chamber synchronous pacemaker.
- Pace in ventricle and sense in ventricle.
- Ventricular pacing inhibited in response to sensing ventricular activity.
- VDD—Ventricular dual chamber synchronous pacemaker.
- Pace in ventricle and sense in both atrial and ventricle.
- Ventricular pacing is triggered or inhibited in response to sensed atrial and ventricular activity.
- Sensed atrial activity will trigger ventricle pacing.However, a sensed ventricular activity will inhibit ventricular pacing.
- DVI—Dual chamber ventricle pacing sequential pacemaker.
- Pace in atrium and ventricle and sense in ventricle.
- Sensed atrial activity triggers a sequence of atrial stimulation, adjustable PR
- interval, and ventricular stimulation.
- Ventricle pacing is inhibited in response to ventricular activity.
- DDD—Dual chamber pacemaker.
- Pace in both atrium and ventricle, sense in both atrium and ventricle.
- Atrium and ventricle pacing inhibited in response to sensing atrial or ventricular activity.
- Atrium pacing inhibited in response to atriumactivity.
- Trigger ventricular pacing if atria activity is sensed unlessventricular activity is sensed.
- VVIRV—Biventricular rate-responsive pacemaker.
- Pace in ventricle, sense in ventricle. Ventricular pacing inhibited in
- response to sensing ventricular activity.
- Automatic rate modulation isenabled and employs multisite ventricular pacing.
- Other rate-responsive pacemakers include AAIR, DDIR, and DDDR.
Types of pacemakers (depends on chamber)
Single Chamber Pacing
- They have only one insulated lead to carry impulses to the heart.
- Either attached to the right atrium or to the right ventricle.
Dual Chamber Pacing
- They have two leads that help the heart to function in a natural way depending on the human body activity.
- One attached to the right atrium and the other to right ventricle.
Triple (Biventricular) Chamber Pacing
- They have three leads and specially designed for patients who have very weak heart muscles; hence they help in the contraction of the heart to produce a normal heart beat.
- One to the right atrium and right ventricle and the other to the left ventricle.